Bennett, A. J. and Harrison, R. G.
ORCID: https://orcid.org/0000-0003-0693-347X
(2006)
In situ calibration of atmospheric air conductivity instruments.
Review of Scientific Instruments, 77 (1).
016103.
ISSN 0034-6748
doi: 10.1063/1.2162459
Abstract/Summary
A new method of measuring the total conductivity of atmospheric air is described. It depends on determination of the electrical relaxation time of a horizontal wire, mounted between two insulators, which is initially grounded and then allowed to charge freely. The total air conductivity derived is compared with that from an ion mobility spectrometer. Results from the two techniques agreed to within 1.2 fS m(-1). (c) 2006 American Institute of Physics.
Altmetric Badge
| Item Type | Article |
| URI | https://reading-clone.eprints-hosting.org/id/eprint/5125 |
| Identification Number/DOI | 10.1063/1.2162459 |
| Refereed | Yes |
| Divisions | Science > School of Mathematical, Physical and Computational Sciences > Department of Meteorology |
| Publisher | American Institute of Physics |
| Publisher Statement | Copyright 2006 American Institute of Physics. This article may be downloaded for personal use only. Any other use requires prior permission of the author and the American Institute of Physics. The following article appeared in Review of Scientific Instruments, 77 (1). 016103 and may be found at DOI: 10.1063/1.2162459. |
| Download/View statistics | View download statistics for this item |
University Staff: Request a correction | Centaur Editors: Update this record