Thermopile radiometer signal conditioning for surface atmospheric radiation measurements

Full text not archived in this repository.

Please see our End User Agreement.

It is advisable to refer to the publisher's version if you intend to cite from this work. See Guidance on citing.

Add to AnyAdd to TwitterAdd to FacebookAdd to LinkedinAdd to PinterestAdd to Email

Harrison, R. G. orcid id iconORCID: https://orcid.org/0000-0003-0693-347X and Knight, J. R. (2006) Thermopile radiometer signal conditioning for surface atmospheric radiation measurements. Review Of Scientific Instruments, 77 (11). p. 116105.

Abstract/Summary

A highly stable microvolt amplifier for use with atmospheric broadband thermopile radiometers is described. The amplifier has a nominal gain of 500, for bipolar input signals in the range +/- 10 mV from a floating source. The noise level at the input is less than 5 mu V (at 100 k Omega input impedance), permitting instantaneous diffuse solar radiation measurements to 0.5 W m(-2) resolution with 12 bit analog to digital conversion. The temperature stability of gain is better than 5 ppm/degrees C (-4 to 20 degrees C). Averaged over a decade of use, the long term drift of the amplifier gain is less than similar to 0.02%/yr. As well as radiometers measuring solar and terrestrial radiations, the amplifier has also been successfully used with low level signals from thermocouples and ground heat flux plates.

Additional Information 116105
Item Type Article
URI https://reading-clone.eprints-hosting.org/id/eprint/5121
Divisions Science > School of Mathematical, Physical and Computational Sciences > Department of Meteorology
Uncontrolled Keywords PYRANOMETER
Additional Information 116105
Download/View statistics View download statistics for this item

University Staff: Request a correction | Centaur Editors: Update this record

Search Google Scholar